Detail výsledku

CPW-fed Double Triangular Slot Antenna for Biomedical applications

HEBELKA, V.; RAIDA, Z. CPW-fed Double Triangular Slot Antenna for Biomedical applications. In 13th Conference on Microwave Techniques COMITE 2013. 2013. p. 43-45. ISBN: 978-1-4673-5513-1.
Typ
článek ve sborníku konference
Jazyk
angličtina
Autoři
Hebelka Vladimír, Ing., Ph.D., REL-SIX (FEKT), UREL (FEKT)
Raida Zbyněk, prof. Dr. Ing., UREL (FEKT)
Abstrakt

In the paper, the design of a coplanar waveguide (CPW) fed double triangular ring slot antenna is studied. The antenna was designed for a wireless body centric communication between a health monitor device and a sensor. The communication link is designed to operate in the ISM frequency band 5.725 GHz to 5.875 GHz (ITU-R 5.138, 5.150 and 5.280). The presented study shows properties of the antenna in various distances from human body models. Attention is turned to impedance matching and radiation pattern of the antenna. The antenna was designed considering a single- layer technology and the Arlon 25N substrate. The prototype of the antenna was fed by a 50 Ohm coaxial cable.

Klíčová slova

Slot antenna, biomedical applications

Rok
2013
Strany
43–45
Sborník
13th Conference on Microwave Techniques COMITE 2013
Konference
13th Conference on Microwave Techniques COMITE 2013
ISBN
978-1-4673-5513-1
BibTeX
@inproceedings{BUT99331,
  author="Vladimír {Hebelka} and Zbyněk {Raida}",
  title="CPW-fed Double Triangular Slot Antenna for Biomedical applications",
  booktitle="13th Conference on Microwave Techniques COMITE 2013",
  year="2013",
  pages="43--45",
  isbn="978-1-4673-5513-1"
}
Pracoviště
Nahoru